Mechanical

RBR data loggers are available in two casings: plastic using the Delrin acetal polymer, or titanium. The depth rating of the plastic housing is 740m; and that of the titanium housing is 8,500m (10,500m for the 1050 and 2050 series cases). Loggers with inductive conductivity cells are limited to 6,600m depth, and with freshwater conductivity cells to 2,000m Other sensors may also restrict the operating depth -- see the above table. Both ends of all XR-series loggers utilize an integral pressure release for safe opening. The O-ring seals at both ends are situated above the threads to ensure a safety pressure release before the end-cap threads fully disengage from the body.

The battery carousel and data port are easily accessed from the non-sensor end. One can access the data port without having to remove the battery carousel. only the end cap needs to be removed. Download time is ~150,000 samples/minute. Batteries are easily changed; simply untwist the end cap and then unscrew the mounting bolt of the carousel.

[CE]The TR-1050 has CE marking EMC Directive 89/336/EEC. This EMC Directive governs conformity with regulations concerning electromagnetic emissions and immunity to electromagmnetic emissions.